Specifications

7 Theory of Operation and Specifications
1006-0939-000 7-23
7-23
Ventilation operating specifications
Pneumatics
Fresh gas
compensation
Pressure
Volume
Gas Source Anesthesia System
Gas Composition Medical Air or O
2
Nominal Supply Pressure 350 kPa
Pressure Range at Inlet 240 to 700 kPa
Peak Gas Flow 120 L/m at 240 kPa, 0.75 seconds
Continuous Gas Flow 80 L/m at 240 kPa
Flow valve range 1 to 120 L/min at 240kPa.
Flow Compensation Range 200 mL/min to 15 L/min
Gas Composition O
2
, N
2
O, N
2
Air, Heliox, CO
2
Anesthetic
Agents
Patient airway pressure range -20 to +120 cmH
2
O +/-2 cmH
2
O
High pressure alarm set range 12 to 100 cmH
2
O, 1 cm increment
Sustained pressure alarm
range
6 to 30 cmH
2
O, 1 cm increment
Display range -20 to 120 cmH
2
O
Tidal volume display range 0 to 9999 mL, 1 mL resolution
Setting range 20 to 1500 mL
Minute volume 0.0 to 99.9 liters, 0.1 liter resolution
Breath rate 4 to 100 bpm (breaths per minute),
1 bpm resolution
Volume sensor type Variable flow orifice