Specifications

Aestiva
3-12 1006-0939-000
3-12
w CC
CC
AA
AA
UU
UU
TT
TT
II
II
OO
OO
NN
NN
No gasket or more than one gasket can cause a leak.
9. Install a new gasket.
10.Align the cylinder post with the index pins.
11.Close the yoke gate and tighten the tee handle.
12.Install a cylinder plug and gasket in all empty cylinder yokes.
13.Do a high-pressure leak test:
Disconnect pipeline supplies.
Disconnect all equipment from the pneumatic outlet and turn Off the
auxiliary flowmeter.
Set the system switch to Standby.
Open the cylinder.
Record the cylinder pressure.
Close the cylinder.
If the cylinder pressure decreases more than 690 kPa in one minute,
there is a leak.
To repair a leak:
Install a new cylinder gasket and tighten the tee handle.
Do this step again. If the leak continues, do not use the system.
ww
ww
WW
WW
AA
AA
RR
RR
NN
NN
II
II
NN
NN
GG
GG
Do not leave gas cylinder valves open if the pipeline supply is in use. Cylinder supplies
could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
DIN connections
1. Close the cylinder valve on the cylinder to be replaced.
2. Loosen the adapter and remove the cylinder.
3. Remove the cap from the cylinder valve on the new cylinder.
4. Point the cylinder outlet away from all items that can be damaged by a
release of high-pressure gas.
AA.96.127
AA.96.129