Specifications

4 User Maintenance
1006-0939-000 4-17
4-17
If the result for all the bellows tests was “passed,” install it in the system.
Before you use the system, complete the preoperative test procedure. Refer to
"Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
Step 5
The bellows must not fall
more than 100 mL/min.
If it does:
The port is not tightly
sealed.
The bellows or the
pressure relief valve is
not correctly installed.
Parts are damaged.