Data Sheet

Specifications subject to change without notice
ATM 065T-F IT / Medical 60 W 65 W
ATM 065P ATM 065A
Features
Suitable for IT / Medical Appliances
Safety: IEC / EN 60601-1 edition 3.1
EMC: IEC / EN 60601-1-2 edition 4.0
Means of Protection: 2 x MOPP
Touch Current: < 100 µA
Efficiency: CEC, DoE Level VI, Energy Star, ErP Stage 2
IEC / EN 60950-1 approval
No Load Power Consumption Less Than 0.15 W
Low Ripple & Noise
1 Year Warranty
Marks / Approvals
Specifications
OUTPUT
MODEL
Voltage
Min. Load
Max. Load
Max. Power
Ripple & Noise
ATM 065T-F120
12 V
0.0 A
5.00 A
60 W
2% Vp-p max.
ATM 065T-F150
15 V
0.0 A
4.34 A
65 W
2% Vp-p max.
ATM 065T-F180
18 V
0.0 A
3.62 A
65 W
2% Vp-p max.
ATM 065T-F190
19 V
0.0 A
3.43 A
65 W
2% Vp-p max.
ATM 065T-F240
24 V
0.0 A
2.71 A
65 W
2% Vp-p max.
ATM 065T-F480
48 V
0.0 A
1.36 A
65 W
2% Vp-p max.
INPUT
100-240VAC
50-60Hz
1.6A-0.7A
Internal Primary Current Fuse
PROTECTION
Yes
Yes
Yes
< 100 µA
4000VAC Primary Secondary
OTHERS
Switching Flyback
0.5 ms for 50% Load Change Typical
10 ms @ full load
Environment
Operating: 0 °C 50 °C (Storage: -20 °C 80 °C)
Operating: 20 % RH 80 % RH (Storage: 10 % RH 90 % RH)
MTBF
300,000 calculated hours at 25 °C, by Telcordia SR-332
SAFETY
ANSI/AAMI/IEC/EN ES60601-1 (edition 3.1); IEC 60950-1 approval
EMI
Compliance to EN 55011 Class B
EMS
Compliance to IEC 60601-1-2
Dimensions
102 mm × 51 mm × 31 mm (4 x 2“ x 1.2“ inch)

Summary of content (1 pages)