Data Sheet

Specifications subject to change without notice
ATM 120T Medical 108 W 120 W
Features
ANSI/AAMI/IEC/EN ES60601-1 (Edition 3.1)
EMC: 60601-1-2 (Edition 4.0)
IEC / EN 60950-1 Approval
Means Of Protection: 2x MOPP
Touch Current: < 100 µA
Regulated Output With Low Ripple Noise
1 Year Warranty
Marks / Approvals
Specifications
OUTPUT
MODEL
Voltage
Min. Load
Max. Load
Max. Power
Connector (mm)
ATM 120T-P/A120
12 V
0.00 A
9.00 A
108 W
4-pin Mini-DIN
ATM 120T-P/A150
15 V
0.00 A
8.00 A
120 W
4-pin Mini-DIN
ATM 120T-P/A180
18 V
0.00 A
6.67 A
120 W
2.1 x 5.5 x 11
ATM 120T-P/A190
19 V
0.00 A
6.32 A
120 W
2.1 x 5.5 x 11
ATM 120T-P/A200
20 V
0.00 A
6.00 A
120 W
2.1 x 5.5 x 11
ATM 120T-P/A240
24 V
0.00 A
5.00 A
120 W
2.1 x 5.5 x 11
ATM 120T-P/A480
48 V
0.00 A
2.50 A
120 W
2.1 x 5.5 x 11
INPUT
Voltage
AC 100-240V
Frequency
50-60Hz
Current (rms max)
1.4A-0.5A
Protection
Internal Primary Current Fuse
PROTECTION
Over Current
Yes
Overvoltage
Yes
Short Circuit
Yes
Over Temperature
Yes
Touch Current
< 100 µA
OTHERS
Topology
Switching Flyback
Dielectric Withstand
4000VAC Primary Secondary
Efficiency
DoE Level VI; Energy Star; ErP Stage 2 Certified
Ripple
2% Vp-p Max. for Output Voltage @ Full Load
Transient Response
0.5 ms for 50% Load Change Typical
Hold-Up Time
10 ms @ Full Load
Load Regulation
±5% Typical
Environment
Temperature
Operating: 0°C 40 °C (Storage: -20°C 80°C)
Humidity
Operating: 20 80 % RH (Storage: 10 90 % RH)
MTBF
SAFETY
EMI
EMS
Case Dimensions
ATM 120T-P ATM 120T-A
IEC 60320 C14 IEC 60320 C8