Users Manual Part 1

The Remunity Pump for Remoduli
(treprostinil) Injection may affect nearby
electrical and electronic devices, including
medical devices. This interference could cause
these devices to operate abnormally or stop
functioning. If nearby equipment looks like it is
being affected by this system, move the Pump,
Remote, or Pump battery charger away from
these machines.
Do not use the system outside of the
environmental conditions listed in the user
guide. Doing so may cause the device to
operate outside of its published accuracy
which can result in harm.
Cell phones, Bluetooth devices, RFID readers,
and Wi-Fi equipment may affect the
communication between the Pump and Remote
or interrupt operation of the Pump. Keep the
Pump and Remote at least 0.8 m [3 ft] away
from these items. See Table 28 on page223
and See Table 30 on page226 for more
information.
Metal detectors may affect pump accuracy or
trigger an alarm. Hand held or walk through
metal detectors may be used near the
Remunity Pump for Remoduli
(treprostinil) Injection, but avoid prolonged
exposure to them. See Table 30 on page226
for more information.
Retail anti-theft detectors may affect pump
accuracy or trigger an alarm. Step through
retail anti-theft detectors at a normal pace, but
avoid standing in them.
See Table 30 on
page226 for more information.
The Remunity Pump for Remoduli
(treprostinil) Injection is MR Unsafe. Remove
the pump before entering an MRI scan room and
do not bring the remote into the MRI scan room.
Contact with or being in proximity to an MRI
scanner can cause the pump and remote to
move or lead to electric shocks and may result
in severe injury.
24 DKPI-09153-001-Rev. 1.4