Cassette Change Warnings
Only Remunity cassettes may be used with
the Remunity Pump for Remoduli
(treprostinil) Injection. Failure to use Remunity
cassettes can lead to harm.
Damaging the portions of the pump exposed
during the cassette and battery changing
process can affect pumping accuracy which
can lead to harm.
Delivery errors may result in adverse medical
events (including serious injury) if the
cassette, attached tubing, and infusion site
catheter are improperly primed.
Contamination to the portions of the pump
exposed during the cassette and battery
change process by dirt, lubricants or liquids
can affect pumping accuracy which can lead to
harm.
Do not connect the tubing to an installed
catheter before the pump completes all self
tests. Doing so may cause over delivery of
medication resulting in harm.
Do not leave the cassette and infusion set
connected to an inserted catheter during
priming operations. Leaving the infusion set
connected during priming operations can lead
to the unintended delivery of medication which
can lead to harm.
Do not connect the cassette to the catheter if
the cassette is not connected to the pump.
Connecting the cassette to the catheter
without the pump connected can lead to the
unintended delivery of Remodulin which can
lead to harm.
Failure to prime the infusion set tubing can
lead to a delay in the delivery of medication,
which can lead to harm.
DKPI-09153-001-Rev. 1.4 25