Risk of fire and burns. Do not open, crush, heat
above 140 °F (60 °C), or incinerate the pump
battery or remote. Doing so can lead to fire or
rapid spreading of fire resulting in harm.
This system supports flow rates between
16µL/h - 225 µL/h. If your flow rate is outside
this range please discuss with your physician.
Replace pump batteries after three months
use. Failure to replace pump batteries after
three months of use may lead to Cassette
Problem alarms.
Always have the following extra supplies with
you if planning to be away from home, even for
a short period of time:
Spare Pump and Remote
Extra non-expired Pharmacy-Filled Cassette
packages (enough for the number of days you
will be traveling) in their unopened packaging.
Extra pump battery with full charge in remote
battery bay.
This User Guide.
Your medication.
Your preferred Infusion Set.
See "Use only
one of the following Infusion Sets" on page33.