Users Manual Part 3

Cassette Change
Stop Delivery Without the
Remote
1. Disconnect the
infusion set tubing
from the catheter.
Refer to the
instructions provided
by your infusion set's
manufacturer for
guidance.
You should consult with your specialty pharmacy
about how often to change the catheter.
Do not remove the pump from the used
cassette until after you have disconnected the
infusion set tubing from your body.
Disconnecting the pump before you
disconnect the infusion set tubing from your
body could lead to unintended medication
delivery.
Disconnect Cassette from
Pump and Dispose without
Remote
To avoid unintended medication delivery,
disconnect from catheter before detaching the
cassette from the pump.
1. Holding the
cassette still,
rotate the pump
counterclockwise
until the pump-
bump is aligned
with the Unlock
symbol.
DKPI-09153-001-Rev. 1.4 85