Alarms and Attention Alarms
Cassette Removed Alarm
When the pump detects that the cassette is
removed without being commanded to stop
delivery, a Cassette Removed Alarm displays.
The top banner of the screen will flash and the
remote will beep.
Do not remove the pump from the used
cassette until after you have disconnected the
infusion set tubing from your body.
Disconnecting the pump before you
disconnect the infusion set tubing from your
body could lead to unintended medication
delivery.
1. Press to silence the alarm.
2. Disconnect infusion set tubing from catheter.
3. Press to acknowledge the alarm.
4. Press to acknowledge the Delivery
Stopped Attention Alarm.
5. Disconnect the cassette from the pump.
6. Remove and reinsert the pump battery into the
pump.
7. Listen for the confirmation beep on the pump.
8. Disregard the Fill new cassette with X.X mL,
prime, and install message if you want to use
your existing cassette.
a. Reconnect the pump to the used cassette.
DKPI-09153-001-Rev. 1.4 151