Users Manual Part 6

a. Disregard the message on the remote telling
you to fill a new cassette with medication,
prime and install. You will not need to do
that since the Occlusion was a kink.
b. Reconnect the pump to the used cassette.
c. Wait for the Self Test to complete.
d. Reattach the infusion set tubing to the
catheter and press and hold the blue
pump button or the
on the remote to
restart delivery.
e. Select No when the New Cassette?
question appears.
f. Route the infusion set tubing to avoid any
sources of kinks that might have introduced
the occlusion that triggered the alarm.
9. If you did not find a kink in the infusion set
tubing you will need to replace the infusion set
tubing as well as the cassette.
10. Dispose of the used cassette and infusion set
tubing.
11. Obtain a new
disposable kit
and perform all
the Cassette
Change steps
starting from
"Replace Pump
Battery" on
page87.
Alarms and Attention Alarms
156 DKPI-09153-001-Rev. 1.4