Instructions / Assembly
The operating instructions for the model BD-2100 are divided
into two sections. Section A is a step-by-step guide to the
different features and basic use of your moisture meter.
Section B contains information on how the meter is used for
testing different materials.
Please read the instructions carefully. If you require additional
information, please contact Delmhorst for assistance.
SECTION A
BEFORE YOU BEGIN
BUTTON FUNCTIONS
! Read Button - This button reads the Percent Moisture
Content value (%MC), in the user-selected scale.
@ Calibration Check Button - This button checks the meter
calibration. It also displays the average of up to 100
readings; displays the highest stored reading; erases the
readings.
# Scale Button - This button sets the display scale to #1
(Wood), #2 (Plaster/Concrete reference scale), and #3
(Gypsum). It also acts as a scroll key, depending on the
function.
$ Set-Point Button - This button programs the set-point
value. A buzzer will alert you if the meter reads higher
than the selected value. It also acts as a scroll key,
depending on the function.
CHECK CALIBRATION
hPress the read button ! and calibration check button @
simultaneously. Meter is in calibration if it displays 12%
(+/- 0.2). Make sure the pins are not in contact with anything
when checking the calibration.
If you check the calibration and the meter does not display
12%, it is likely an indication of a low battery. If this occurs,
change the battery immediately. Continued use with a low
battery may cause the meter to go out of calibration. If you
have a fresh battery and the instrument still does not indicate
a proper calibration, return it to DELMHORST for service.
See Service for Your Meter section.
When the battery is removed and then reconnected, the
meter displays its software version for one second and then
turns itself off. After replacing the battery, you must reset
the meter as described in Resetting the Meter section.