Instructions / Assembly
5
INFORMATION ABOUT YOUR
READINGS
The meter will accumulate up to 100 readings. After all 100
readings are “stored” it will not add new readings until the
memory has been cleared. It will also continue to display the
average of all 100 readings as a reminder that the memory is full.
hTo add a reading to the sum of all the previously stored
readings, release the read button
! within 2 seconds.
When taking and storing readings of a specific material, be
sure to “clear” the meter before moving on to the next scale
if you do not want to group all of the readings together.
Readings below 6% will be displayed as -5.8 and readings above
40% will be displayed as a flashing 40.0. Neither of these types
of readings will be added to your accumulated readings.
TO CHECK ACCUMULATED
READINGS
This feature allows you to view the total number of all
accumulated readings for the given material you have chosen,
the average of those readings, and the highest stored reading.
hTo view the readings press and release the calibration
check button
@. First the meter displays the number of
accumulated readings for one second, then the average of
those readings for two seconds. Then it displays the highest
stored reading for two seconds. The total “cycle” time is
five seconds.
hTo erase readings hold the calibration check button
@ for
5 seconds. All accumulated readings will be erased and the
meter will display “0”.
hTo keep the accumulated readings in memory, release the
calibration check button
@ before the meter finishes the
above cycle.
TO RESET METER
hPress and release the calibration check button @.
hWithin one second press the scale button
#.
hThe meter will reset itself the default setting of Scale #1
(wood) and 15% set-point, and clear all the readings stored
in memory.