REV. 1.1 10/06/2016 HD21ABE HD21ABE17 IAQ MONITOR ENGLISH The quality level of our instruments is the result of the continuous development of the product. This may produce some differences between the information written in this manual and the instrument you have purchased. We cannot completely exclude errors in the manual, for which we apologize. The data, images and descriptions included in this manual cannot be legally asserted. We reserve the right to make changes and corrections with no prior notice.
HD21ABE Indoor Air Quality Monitor - 2 -
HD21ABE 1. Backlit graphic display. 2. ESC key: It allows to exit from the menu or, in case of a submenu, to exit from the current level display. 3. Navigation key ▲: It allows navigation through the menus. During normal operation, it is used to select the resetting of the statistical data and to scroll the displayed quantities upwards. 4. Navigation key ◄/Func: It allows navigation through the menus. In normal view, it allows to display the statistical data: maximum, minimum, and average. 5.
HD21ABE17 Indoor Air Quality Monitor - 4 -
HD21ABE17 1. Backlit graphic display. 2. ESC key: It allows to exit from the menu or, in case of a submenu, to exit from the current level display. 3. Navigation key ▲: It allows navigation through the menus. During normal operation, it is used to select the resetting of the statistical data and to scroll the displayed quantities upwards. 4. Navigation key ◄/Func: It allows navigation through the menus. In normal view, it allows to display the statistical data: maximum, minimum, and average. 5.
TABLE OF CONTENTS 1. GENERAL CHARACTERISTICS ...................................................................................................................... 7 2. THE USER INTERFACE .................................................................................................................................... 8 2.1 THE HD21ABE DISPLAY ....................................................................................................................................... 8 2.2 THE HD21ABE17 DISPLAY .
1. GENERAL CHARACTERISTICS HD21ABE and HD21ABE17 IAQ Monitors are bench-top/portable instruments manufactured by Delta Ohm for the analysis of indoor air quality (IAQ, Indoor Air Quality). The instruments simultaneously measure the parameters: CO2 Carbon Dioxide, CO Carbon Monoxide and Atmospheric Pressure. The HD21ABE17 instrument also measures Temperature and Relative Humidity, and it calculates Dew Point, Wet Bulb Temperature, Absolute Humidity, Mixing Ratio and Enthalpy.
2. THE USER INTERFACE The user interface consist of a backlit LCD graphic display, and the power-on and setting keys. When battery powered, and not pressing any key, the backlight turns off after about 1 minute. To turn it back on, press any key. When using an external power supply, the backlight is always on. Turn the instrument on and off with the ON/OFF key. When you turn the instrument on, the logo and model will be displayed for a few seconds, and then the main display.
2.2 THE HD21ABE17 DISPLAY 1 2 3 1. 2. 3. 4. HD21ABE17 2010/02/10 08:00:00 Carbon Dioxide CO2 ppm CO 0 ppm Patm 1000 hPa RH 25.0 % T 17.0 °C 4 600 Battery’s charge status and instrument code. In case the logging function is on, this line shows the current logging number and the time elapsed from logging start. Main quantity (in this case, CO2 Carbon Dioxide). Display of all other quantities. Current date and time.
2.3 THE KEYBOARD The keys on the instrument perform the following functions: ON-OFF/AUTO-OFF key It turns the instrument on and off. When turning the instrument on, the first screen will be displayed. After few seconds the measured quantities will be displayed. + Auto Power Off The instrument has an AutoPowerOff function that automatically turns the instrument off after about 8 minutes if no key is pressed.
Key ◄/FUNC ◄ allows navigation through the menus. FUNC: In normal view, it allows to select the statistical data: maximum, minimum, and average. Key ▲ It allows navigation through the menus. During normal operation, it is used to select the resetting of the statistical data and to scroll the displayed quantities. Key ▼ It allows navigation through the menus. During normal operation, it is used to cancel the resetting of the statistical data and to scroll the displayed quantities.
3. OPERATION During turning on, the following message is displayed for about 10 seconds: Model HD21ABE17 Indoor Air Quality Firm.Ver.=01.00 In addition to the Delta Ohm logo the instrument code and the firmware version are displayed.
3.1 “►/UNIT” KEY FOR THE UNIT OF MEASUREMENT It allows navigation through the menus. During normal operation, it changes the unit of measurement of the main quantity (solo HD21ABE17).
Function: AVG 1min RUNNING 00:00:48 Carbon Dioxide CO2 ppm CO 0 ppm Patm 1000 hPa RH 25.0 % T 17.0 °C 600 After 1 minute the instrument emits one long beep and displays the calculated average on the LCD. Function: AVG 1min AVG 1min 00:00:00 Carbon Dioxide CO2 ppm CO 0 ppm Patm 1000 hPa RH 25.0 % T 17.0 °C 602 The calculated average is frozen on the LCD until the ENTER key is pressed. To escape the 1min average calculation during the countdown, press the ESC key.
4. MAIN MENU To access the programming menu press MENU: MAIN MENU 1) Informations 2) Logging 3) Serial 4) Settings 5) Probes Calibration 6) Language ▲▼ select confirm If you do not press any key within 2 minutes, the instrument goes back to the main display. Use the arrows ▲▼ and press ENTER to select an item. To exit the selected item and return to the previous menu, press ESC. To exit immediately from the main menu, press MENU again. 4.1 INFO MENU Enter the main menu by pressing MENU.
To change the ID, press ENTER. Use the ◄► arrows to select the item, and edit it with the ▲▼ arrows. Proceed with the other items, and finally confirm with ENTER. By selecting Info Sensors, the following information on the sensors will be displayed: INFO SENSORS Type= CO2-CO Fw.V0R0 Cal = 2010/02/10 SN = 10010000 INFO SENSORS: Firmware type and version. Calibration date. Serial number of the sensors board. Press ESC to return to the main menu. Press MENU to exit the menu.
4.2 LOGGING MENU Enter the main menu by pressing MENU; • Use the arrows ▲▼ to select Logging; • Press ENTER: The parameter setting submenu for the logging sessions (to be captured) will be displayed. LOGGING MENU 1) 2) 3) 4) 5) Log frequency Auto switch off Start/Stop Log Start Log Erase Log File Manager ▲▼ select confirm 4.2.
4.2.2 AutoPowerOff– AutoPowerOff Mode The Auto switch off item controls the instrument’s automatic turning off during logging, between the capture of a sample and the next one. When the interval is lower than 5 minutes, the instrument will always remain on. With intervals greater than or equal to 5 minutes, it is possible to turn off the instrument between loggings: it will turn on one minute before sampling and will turn off immediately afterwards, thus increasing the battery life.
4.2.3 Start/Stop Log – Automatic Start The logging start and end can be programmed by entering the date and time. Set the logging start date and time using the arrows. Confirm the logging start date and time using ENTER. Then you are asked to set the data to end the recording. Set the logging end date and time using the arrows. Confirm the logging end date and time using ENTER. To enter this setting, proceed as follows. Once you have accessed the LOGGING submenu (previous par.
8. Once both values have been set, a summary will be displayed showing the start and end time of the LOG session. LOGGING MENU SETTED LOG START Date 2010/02/10 10:29:00 END Date 2010/02/10 10:39:00 exit/cancel confirm 9. Press ENTER to confirm or ESC to exit without enabling the automatic start: In both cases, you will return to the LOGGING menu. 10. Press MENU to exit immediately from the main menu.
4.2.4 Cancel Auto-Start Once the LOG session start and end times are set, you can inhibit the session automatic start by using Start Log Erase. Once you have accessed the LOGGING submenu: 1. Use the arrows ▲▼ to select Start Log Erase 2. The LOG session start and end times will be displayed: MENU LOGGING Auto-Start Erase Setted start: 2010/02/10 10:29:00 Setted end: 2010/02/10 10:39:00 Press ▲▼ for Auto-Start Erase confirm 3.
4.2.5 Log File Manager This item allows managing the logs captured: The instrument allows displaying of the captured data files and erasing the entire memory (File Log Erase). The instrument can store up to 64 LOG sessions, progressively numbered from 00 to 63: The session list is arranged on 4 lines and 4 columns. If you have over 16 sessions, using the MEM function key you shift to the next screen. The current page (0, 1, 2 or 3) and the total data pages are displayed in the upper right corner.
See File Log: Selecting this item you can view the logging sessions in the instrument: LOG FILE 0/3 00– 01– 02– 03 04– 05– 06– 07 08– 09– 10– 11 12– 13– 14– 15 START Date: 2010/02/10 08:50:40 Record: 000039 ▲▼◄► select page change 1. Use the arrows ▲▼◄► to select the log, and the MEM key to shift page. 2. Once a file is selected, the acquisition start date and time and the number of samples contained in the file (Rec) are displayed at the bottom of the display. The files are stored in ascending order.
Log lasting (time set for recording) It represents the recording duration: After this set time, the recording is ended. The recording can be stopped earlier by pressing MEM. To disable this function, set the time to 0:00:00. In this case the recording ends by pressing MEM or when the memory is full. LOGGING MENU LOG LASTING h:mm:ss (1h max) 00:00:00 With setting: 00:00:00 Log stop with key MEM ▲▼ set exit Use the arrows to change the set time; the maximum allowed value is 1 hour. Confirm with ENTER.
4.3.1 Print Interval To set the Print Interval, proceed as follows: 1. Use the arrows ▲▼ to select the item; 2. Press ENTER: You will get the following message: COMMUNICATION MENU SERIAL PRINT INTERVAL h:mm:ss (1h max) 0:00:00 ▼▲ set exit confirm 3. Use the arrows ▼▲ to set the value; 4. Press ENTER to confirm and return to the previous page, or press ESC to cancel the change and exit the menu item; 5. Press ESC over and over to exit from the submenus; 6.
4.4.1 Contrast This item of the Settings menu allows to increase or decrease the display contrast. To access the Contrast submenu, proceed as follows: 1. Use the arrows ▲▼ to select Contrast; 2. Press ENTER 3. You will get the following message CONTRAST LCD Contrast set: 012 ▼▲ set exit/cancel 4. Use the arrows ◄► to decrease or increase the contrast; 5. Press ENTER or ESC to return to the main menu; 6. Press MENU to exit immediately from the main menu. 4.4.
4.5 PROBES CALIBRATION The instruments and sensors are calibrated in the factory; no calibration is usually required by the user. However, you can perform a new calibration. You can perform the calibration of the CO (Carbon Monoxide) and CO2 (Carbon Dioxide) sensors. With the HD21ABE17 instrument you can also perform the calibration of the RH (Relative Humidity) sensor. No calibration is allowed for the Temperature sensor (HD21ABE17).
PROBE CALIBRATION CALIBRATION CO2 1) Offset only calib 2) 2 Pts Calibration 3) Restore Fact cal ▼▲ select exit/cancel confirm The CO2 sensor can be calibrated: • • In 1 point: only the sensor offset is corrected, the point can be any value between 0 and 950 ppm. In 2 points: the sensor offset and slope are corrected, the lower point can be any value between 0 and 950 ppm, the upper point can be any value between 1000 ppm and the full scale of the instrument. 1 point calibration: 1.
8. If the nitrogen bottle was used, close the bottle valve, remove the tube from the HD21AB17.9 accessory and remove the accessory from the instrument. Calibrazione in 2 punti: 1. Place the instrument in an environment with known CO2 concentration between 0 and 950 ppm (for ex. in clean air). For 0 ppm calibration with nitrogen bottle (code MINICAN.12A), slide the HD21AB17.
14. Adjust the value by using the arrows ▼▲. 15. Press ENTER on the instrument and wait the time necessary for calibration without changing the working conditions. 16. When the “CALIBRATION GOOD” indication appears, press ESC to return to the calibration menu. Restoring the factory CO2 calibration: In case of erroneous execution of the calibration procedure, it is always possible to get back to the factory calibration by selecting the 3) Restore Fact cal calibration option.
4.5.2 CO Calibration You can calibrate the zero of the CO sensor in clean air (the CO concentration is lower than 0.1 ppm outdoor) or using nitrogen bottles (code MINICAN.12A). Use the arrows ▲▼ to select the 2) Calibration CO item: PROBES CALIBRATION 1) Calibration CO2 2) Calibration CO 3) Calibration RH ▼▲ select exit/cancel confirm Confirm with ENTER.
CO zero calibration with nitrogen bottle (code MINICAN.12A): Using a screwdriver, open the back small door of the instrument. Connect the tube from the MINICAN.12A bottle with the rubber cap on the CO sensor head. Back door CO sensor Use the arrows ▲▼ to select the 2) Calibration CO item: PROBES CALIBRATION 1) Calibration CO2 2) Calibration CO 3) Calibration RH ▼▲ select exit/cancel confirm Confirm with ENTER.
Confirm with ENTER. The following screen will appear: PROBES CALIBRATION CO CALIBRATION CAL ZERO Cal zero 0 ppm exit/cancel confirm • • • • • Wait about 15 minutes before continuing. Supply the gas adjusting the bottle flow meter to get a constant flow from 0.1 to 0.2 l/min. Press ENTER and wait the two minutes necessary for calibration without changing the working conditions. At the end, close the bottle spigot and remove the cap from the CO sensor. Insert the protection grid.
3. Select 2) Set sensitivity and confirm with ENTER. The following screen will appear: PROBES CALIBRATION CALIBRATION CO SENSITIVITY SET Sens CO 50 nA/ppm 0 ppm ▼▲ set exit/cancel confirm 4. Use the arrows ▼▲ to change the CO sensor sensitivity value, and confirm using ENTER.
4.5.3 RH calibration (HD21ABE17 only) This procedure applies to HD21ABE17 instrument. Before starting the calibration operation, you should check if you need a new calibration using the saturated solutions at 75.4%RH and 33%RH: You should proceed with the calibration, only if you find an error in one of the two above calibration points. The calibration procedure removes the previous calibration data. For a correct calibration of the sensor, the first point should be 75%RH and the second point 33%RH.
4. The probe and the saturated solution for this operation should have the same temperature, and therefore should be located in a room with a stable temperature for the entire calibration period. 5. Unscrew the probe protection; screw the M12×1 threaded ring. 6. Should there be any liquid inside the measuring chamber, dry it with some blotting paper. The formation of liquids inside the measuring chamber, does not impair the measurement uncertainty of the solution or measurement. 7.
Relative Humidity Balance values of some saturated salt solutions from 0°C to 100°C Temp. °C 0 Lithium chloride 11.23 ± 0.54 Magnesium chloride 33.66 ± 0.33 Sodium chloride 75.51 ± 0.34 5 11.26 ± 0.47 33.60 ± 0.28 75.65 ± 0.27 10 11.29 ± 0.41 33.47 ± 0.24 75.67 ± 0.22 15 11.30 ± 0.35 33.30 ± 0.21 75.61 ± 0.18 20 11.31 ± 0.31 33.07 ± 0.18 75.47 ± 0.14 25 11.30 ± 0.27 32.78 ± 0.16 75.29 ± 0.12 30 11.28 ± 0.24 32.44 ± 0.14 75.09 ± 0.11 35 11.25 ± 0.22 32.05 ± 0.13 74.87 ± 0.
5. CONNECTION TO A PC HD21ABE and HD21ABE17 are fitted with an USB 2.0 interface. Optionally, you can receive on request the serial connection cable CP23 with mini-USB connector on instrument’s side and USB 2.0 connector on computer’s side. The instruments are supplied with the DeltaLog10 software (from version 0.1.5.3) running in a Windows® operating environment. The software manages the connection, data transfer, graphic presentation, and printing operations of the captured or logged measurements.
Command KE LE K1 K0 K4 K5 KP KQ WC0 WC1 RA Response & & & & & & & & & & Sample print = RL WA# Sample log = & WL# & 0sec 30sec Description Stop the data download Erase stored data Immediate printing of data Stop printing data Start logging data Stop logging data Auto-power-off function = ENABLE Auto-power-off function = DISABLE Setting SELF off Setting SELF on Reading of PRINT interval set and label of the measurements Reading of LOG interval set Setting PRINT interval.
• Some keys are disabled during logging. The following keys are enabled: MEM, MENU, ENTER and ESC. • Pressing the MEM and MENU keys has no effect on the logged data if these keys are pressed after starting the recording, otherwise the following is valid. 5.1.3 Print function Press ENTER to send the measured data directly to the USB port, in real time. The printed data units of measurements are the same as those used on the display. The function is started by pressing ENTER.
6. INSTRUMENT SIGNALS AND FAULTS The following table lists all error indications and information displayed by the instrument and supplied to the user in different operating situations: Display indication Explanation ---- This appears if the sensor relevant to the indicated physical quantity is not present or is faulty. OVFL Overflow appears when the probe detects a value that exceeds the expected measurement range.
7. BATTERY SYMBOL – MAINS POWER SUPPLY The meter is provided with a pack of 4 x 1.2V-2200mA/h Ni-MH rechargeable batteries, placed in the battery compartment. The battery symbol on the display constantly shows the battery charge status. To the extent that batteries have discharged, the symbol "empties". When the charge decreases still further it starts blinking. In this case, batteries should be replaced as soon as possible.
7.2 NOTES FOR THE BATTERIES USE • • • • • • • • • At the first start up, it’s necessary to completely recharge the batteries. The charge time of the batteries package is about 4 hours. The last of the batteries package in measurement working mode is about 8 hours. A new Ni-MH batteries package reaches the maximum of its performance only after being discharged and charged completely again at least twice or three times. The batteries package autonomy depends on the instrument use.
8. INSTRUMENT STORAGE Instrument storage conditions: • Temperature: -25…+65°C. • Humidity: less than 90% RH without condensation. • During storage avoid locations where: • humidity is high; • the instrument may be exposed to direct sunlight; • the instrument may be exposed to a source of high temperature; • the instrument may be exposed to strong vibrations; • the instrument may be exposed to steam, salt or any corrosive gas.
9. TECHNICAL CHARACTERISTICS Instrument Dimensions (Length x Width x Height) 210x90x40 mm (HD21ABE) 300x90x40 mm (HD21ABE17 with probe) 470 g (batteries included) ABS, rubber Backlit, Dot Matrix 160x160 dots, visible area 52x42 mm Weight Materials Display Operating conditions Operating temperature Warehouse temperature Working relative humidity -5 … 50°C -25 … 65°C 0 … 85% RH without condensation Instrument uncertainty ± 1 digit @ 20°C Power Mains adapter (code SWD10) Batteries 12Vdc/1A 4 x 1.
9.1 TECHNICAL DATA OF THE SENSORS CO2 Carbon Dioxide Sensor Measurement range Sensor working range Accuracy Resolution Temperature dependence Response time (T90) Long-term stability NDIR Dual Wavelength 0 … 5000ppm -5 … 50°C ±50ppm+3% of measure 1ppm 0.1%f.s.
10. ORDERING CODES HD21ABE IAQ Monitor datalogger kit. It measures CO, CO2 and atmospheric pressure. Complete with: DeltaLog10 software (version 0.1.5.3 and later) for data download, monitor, and data processing on Personal Computer, 4 x 1.2V NiMH rechargeable batteries, operating manual, case. The cables must be ordered separately. HD21ABE17 IAQ Monitor datalogger kit. It measures CO, CO2, atmospheric pressure, temperature and relative humidity. Complete with: DeltaLog10 software (version 0.1.5.
NOTES - 48 -
NOTES - 49 -
NOTES - 50 -
Dichiarazione di conformità UE EU declaration of conformity Prodotto: Indoor Air Quality Monitor HD21ABE – HD21ABE17 Product: Indoor Air Quality Monitor HD21ABE – HD21ABE17 Accessori: Alimentatore stabilizzato SWD10 Accessories: SWD10 stabilized power supply Fabbricante: Manufacturer: Delta Ohm S.r.l. a socio unico via G. Marconi 5 35030 Caselle di Selvazzano (PD) ITALY La presente dichiarazione di conformità è rilasciata sotto la responsabilità esclusiva del fabbricante.
GU AR AN T E E TERMS OF GUARANTEE All DELTA OHM instruments are subject to accurate testing, and are guaranteed for 24 months from the date of purchase. DELTA OHM will repair or replace free of charge the parts that, within the warranty period, shall be deemed non efficient according to its own judgement. Complete replacement is excluded and no damage claims are accepted. The DELTA OHM guarantee only covers instrument repair.