User Manual

ENGLISH 16
Difficulty removing outer sheath from inner module 1. Check the components for gross debris
2. Hold the handpiece outer sheath securely and twist the
inner module
3. Inspect parts for wear
4. Call Technical Support if necessary
Charging base does not drain liquids Clear any debris from the hole at the bottom of the
charging base.
Handpiece does not charge 1. Clean the charge contacts on the handpiece and
charging base, using one of the approved cleaning
solutions described in section 8.
2. Verify that the power supply is properly connected to
the charging base and that the green LED on back
lights up.
3. Verify that the handpiece is able to properly sit inside
the charging base, and that there are no foreign
obstructions.
4. If still not charging, return the handpiece to Technical
Support for professional battery replacement.
Handpiece does not hold charge 1. Verify that the handpiece properly charges (LEDs scroll
when in the charger).
2. Return the handpiece to Technical Support for
professional battery replacement.
Orange Service Light in handpiece illuminates 1. Handpiece is heating up due to excess ON time or
load. Do not use handpiece for several minutes and
allow it to cool down.
2. If LED still illuminates even after allowing handpiece to
cool down, set aside and call Technical Support.
Power supply overheating Immediately unplug the unit and call Technical Support.
Power cords are frayed or damaged in any way Do not use. Call Technical Support.
NOTE:
For current Technical Support contact information, see “Manufacturer contact“ in the infection control procedures
(section 8).