User Manual

8/17/2005 CavitronRF Combo - DFU Draft 08-01-05 Rev8doc
DRAFT
Page 10 of 18
Insert Tip
Shape and size of tip determines access and adaptation.
Insert Marking
Manufacturer, Date (YYMM),
Frequency, Insert Type (E.g. DENTSPLY 0508 30K FSI-SLI-10S)
Hold the handpiece in an upright position. Activate the Foot Control to bleed any air bubbles that
might be trapped inside the handpiece. Lubricate the O-ring on the insert with water before placing it
into the handpiece. Fully seat insert with a gentle push-twist motion. DO NOT FORCE.
The many styles of Cavitron® 30K™ Ultrasonic Inserts are easily interchangeable for various
procedures and applications. See enclosed booklet for specific information.
6.5 Cavitron® JET Air Polishing Inserts
(Picture from current DFU’s)
Air Polishing Insert Nozzle
Tube-in-tube design delivers precise air/water/powder mixture at point of delivery.
Prophy Powder Delivery Tube
Directs air/water powder flow to nozzle
Insert Rod
Warms lavage for patient comfort.
O-Ring
Provides a watertight seal when the insert is fully seated in the handpiece. O-ring should be replaced when
worn.
The Air Polishing Insert automatically selects the Air Polishing mode when placed in the handpiece.
Lubricate the rubber O-ring on the insert with water before placing it into the handpiece. Place the
insert rod into the insert port of the handpiece. Align the powder delivery tube with the powder
delivery port and gently push into handpiece until fully seated.
6.6 Foot Control Information and Operation
The Foot Control is a two-positioned momentary switch. The first position activates both ultrasonic energy
and lavage at the insert tip for scaling procedures and lavage only (rinse) at the air polishing insert nozzle
for air polishing procedures. The second position (fully depressed footswitch) activates Boost Mode while
scaling and, with a JET insert present in the handpiece, enables air/powder/water slurry for air polishing
procedures. The Boost Mode provides capability to briefly increase the system’s ultrasonic power output
for quick / efficient removal of tenacious deposits using only the foot control without touching the system
base.
Pressing anywhere on the top of the footswitch activates the base unit.
(Picture showing released, 1
st
position, 2
nd
position)
In the event that the batteries do run low and new batteries are not available for replacement, an
auxiliary cord is supplied to provide emergency control.
(Picture to show hook-up)
Section 7: Accessories and User Replaceable Parts
7.1 Accessories
1. AC Power Cord
2. Dual Position Foot Control (Cordless)
3. Auxiliary Footswitch Cable
4. Cavitron® JET-Mate Sterilizable Handpiece
5. Prophy Handpiece Cleaning Wire
6.
Cavitron® Ultrasonic Insert
Formatted: Bullets and Numbering