Page 6 of 18
5.2 Water Line Requirements
A water supply line with user-replaceable filter is supplied with your system. See Section 9.0 System
Care for replacement instructions.
Incoming water supply line pressure to the system must be 20 psi (172kPa minimum) to 40 psi
(275kPa maximum). If your dental water system’s supply line pressure is above 40 psi, install a water
pressure regulator on the water supply line to your Cavitron® RF™ Combination System.
A manual shut-off valve on the dental water system supply line should be used so that the water can be
completely shut-off when the office is unoccupied.
In addition to the water filter supplied, it is recommended that a filter in the dental water system supply
line be installed so that any particulates in the water supply will be trapped before reaching the system.
After the above installations are completed on the dental water supply system, the dental office water
line should be thoroughly flushed prior to connection to the system.
5.3 Air Line Requirements & Recommendations
An air supply line with a user-replaceable filter assembly is supplied with your Cavitron® RF
Combination System. A filter mounting bracket is included for hanging the air filter. The clear bowl
should hang downward allowing for moisture separation and drainage of water from the air filter. See
Section 9.0 System Care for replacement instructions.
Incoming air supply line pressure to the system must be 65 psig (448kPa) to 100 psig (690kPa). If
your office air line pressure is above 100 psig (690kPa), install an air pressure regulator on the supply
line to your Cavitron® RF™ Combination System.
A manual shut-off valve on the office air supply line should be used so that the air line can be
completely shut-off, and the line pressure relieved when the office is unoccupied.
The system must be supplied with clean, dry air to help prevent water condensation from forming in
the air supply line which may cause the system to malfunction. In addition to the air filter supplied
with your system, it is strongly recommended that an air dryer be used on the compressor line
supplying the system.
5.4
Electrical Requirements
Incoming power to the system must be 100 volts AC to 240 volts AC, single phase 50/60 Hz capable
of supplying 1.0 amps.
The system power should be supplied through the AC power cord provided with your system. The
power cord should be attached to an approved hospital-grade AC wall outlet for safe and effective
operation.
5.5
Unpacking the System
Carefully unpack your Cavitron® RF™ Combination System and verify that all components and
accessories are included:
1. Cavitron® RF™ Combination System and factory installed handpiece cable assembly
2. Air Line Assembly (Black) with Filter and Quick Disconnect
3. Water Line Assembly (Blue) with Filter and Quick Disconnect
4. Additional Water Line Filter
5.
Detachable AC Power Cord
6. Cordless Foot Control Assembly
7. “AA” Batteries
8. Auxiliary Cable for cordless Footswitch
9. Cavitron® JET Air Polishing Nozzle with cleaning tool
10.
Jet-Mate Detachable Sterilizable Handpiece
11. Prophy Handpiece Cleaning Wire
12.
Cavitron® Ultrasonic Inserts (quantity optional)
13. Efficiency Indicator for Cavitron Inserts
14. JetShield™ Aerosol Reduction Device Kit
15.
Literature Packet
16. PROPHY-JET® Sodium Bicarbonate Prophy Powder
17. JET-Fresh® Aluminum Trihydroxide Prophy Powder
5.6 System Installation