Specifications

8 Synthes Colibri Instructions for Use
Using the Colibri
Charging, storing and using batteries
Charging
The batteries should always be charged before use.
Only use chargers recommended by Synthes to charge the
batteries. Using a charger that does not originate from
Synthes can damage the battery.
Charge the batteries within an ambient temperature range
of 0°C to 40°C.
Storage
When the battery is not being used, store it in the appro-
priate Synthes charger. This will keep it from discharging,
and the battery will be fully charged and ready to use.
Before first use or after storing the battery outside of the
charger for more than one month, up to five charging/
discharging cycles may be necessary for the battery to be
fully charged.
The charging station should always be turned on when a
battery is in the charging base. This ensures availability and
prevents discharge.
Use
Only insert the battery pack directly before using the
Colibri. This saves battery energy and prevents having to
change it during surgery.
Never open the battery casing during surgery.
Only use batteries for the indicated purpose.
The Colibri 14.4 VDC battery (532.033) may not be used
in oscillating mode since this could damage the motor
and gears. For this mode, always use the standard 12 VDC
battery (532.003).
Only use the 14.4 VDC Colibri battery with the attach-
ments marked with a * on pages 16 and 17.
Do not transport or store the batteries together with mate-
rials that conduct electricity and can cause a short. This
can damage the battery and generate heat which can
cause burns.
Never expose batteries to temperatures above 60°C.
Do not apply force to the batteries, and do not let them
fall.
Never use damaged batteries; they can damage the power
tool.
Follow the additional information in the section “Care and
Maintenance“ starting on page 21 as well as the Instruc-
tions for Use of the Synthes Universal Battery Charger
(036.000.725).