Use Instructions
20
EMC INFORMATION
The NIRA Skincare Laser needs special precauons regarding EMC and needs to be installed according to EMC informaon. The d evice contains hardware capable of Bluetooth Low
Energy (BLE) communicaon. This hardware can receive and transmit RF energy in 2400 —2483.5 MHz frequency range with DSSS modulaon. It is compliant with CFR Title 47, FCC part
15.249 and ETSI EN 300-400 standards, and equivalent isotropically radiated power is less than 10 dBm (10mW).
Cauon: Mobile RF communicaons equipment can aect the NIRA Skincare Laser device.
Warnings: The use of accessories and cables other than those specied in this document may result in increased emissions or decreased immunity of the NIRA Skincare Laser.
Warnings: The NIRA Skincare Laser should not be used adjacent to or stacked with other equipment treatments.
Electromagnec Emissions Declaraon
The NIRA Skincare Laser is intended for use in the electromagnec environment specied below. The customer or the user of the NIRA Skincare Laser should assure that it is
used in such an environment.
RF emissions CISPR 11 Group 1
The NIRA Skincare Laser uses RF energy only for its internal funcon. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B
The NIRA Skincare Laser is suitable for use in all establishments, including domesc establishments and
those directly connected to the public low-voltage power supply network that supplies buildings used for
domesc purposes.
Harmonic emissions IEC 61000-3-2 Complies
Voltage uctuaons/ icker emissions IEC
61000-3-3
Complies