INSTRUCTION MANUAL Detcon Model 974-071170-BCR FLOW SAMPLE FLOW BYPASS FLOW www.detcon.com BALLAST GAS DETECTION MONITOR SAMPLE SAMPLE CALIBRATE BYPASS 4 Channel Sequential Ballast Monitor DETCON, Inc. 3200 Research Forest Dr, The Woodlands, Texas 77387 Ph.281.367.4100 / Fax 281.298.2868 Hwww.detcon.com Oct. 15, 2009 • Document # 3494 • Revision 0.
Ballast Monitor I.M. This page left intentionally blank Ballast Monitor I.M.
Ballast Monitor I.M. Table of Contents 1. 2. 3. Introduction ..................................................................................................................................................1 Installation ....................................................................................................................................................1 Start Up..................................................................................................................................
Ballast Monitor I.M. This page left intentionally blank Shipping Address: 3200 A-1 Research Forest Dr., The Woodlands Texas 77381 Mailing Address: P.O. Box 8067, The Woodlands Texas 77387-8067 Phone: 888.367.4286, 281.367.4100 • Fax: 281.292.2860 • Hwww.detcon.comH • Ballast Monitor I.M.
Ballast Monitor I.M. 1. Introduction The Detcon Ballast Monitor is a sequential 4 channel gas monitor unit. The sensor assembly, discussed in the attached sensor manual, is specified at time of order as to gas and range of detection and other options. The NEMA 7X rating is by definition explosion proof and therefore is suitable for outdoor location in electrically hazardous environments. The enclosure houses a 24 VDC power supply, flow pump and flow fault detection circuit.
Ballast Monitor I.M. 2. Install four lengths of tubing from the desired sample points to the appropriate ‘Sample Port’ inlet on the Ballast Monitor. There are four input ‘Sample’ ports that should be connected to for separate sample points. Note: When detecting for NH3, CL2, CLO2, or HCL care must be taken to use a minimum amount of stainless steel tubing and connections because stainless steel readily absorbs these gasses and will cause significant error in the detection process.
Ballast Monitor I.M. Figure 3 Flow Fault PCA Adjusting the Flow Fault PCA trip point: 1. Completely close the Bypass Rotameter to ensure that it does not affect the reading of the Flow Rotameter. 2. Use the Sample Flow Rotameter to set the air flow at the point where the fault condition should activate. 3. Turn the flow fault adjust potentiometer (PT1 in Figure 3) clockwise until the flow fault LED (D2) illuminates and the associated relays fire.
Ballast Monitor I.M. 4. Calibration The Ballast Monitor and the attached sensor are calibrated prior to shipment and should require minimal adjustment at the time of installation. However, it is recommended that a complete calibration is performed on a periodic basis to assure optimum unit performance. Refer to the sensor’s instruction manual for gas type, range, and sensor specific calibration instructions.
Ballast Monitor I.M. 5. Spare Parts Part# 348-900000-000 350-02QDSW-024 350-601403-005 351-900000-024 360-205420-024 500-005151-002 823-600000-000 975-041400-000 976-041400-000 Description 0-1 Lpm Rotameter with Valve 24VDC Solenoid 2 Position Valve 0-5 Lpm Rotameter with Valve Pump, Dual Diaphragm, 120VAC 24VDC 40Watt power supply Flow fault circuit assembly Genie Membrane Moisture Filter Model RL-4 Addressable Relay Module Model DA-4 Four Channel Input Module 6. Specifications 6.
Ballast Monitor I.M. 8. Flow Diagram Sample 4 Sample 3 Sample 2 Sample 1 Solenoid (1) Genie Membrane Solenoid (2) Solenoid (3) Solenoid (4) Flow Fault Pump 10Psip Pressure Relief Power Supply AC Power Calibrate Vent RS-485 RL4 DA4 Sample FP-524 Roto Meter 3-Way Valve Roto Meter Vent Figure 5 Flow Diagram Ballast Monitor Instruction Manual Rev. 0.
Ballast Monitor I.M. 9.
Ballast Monitor I.M. Appendix A Revision History Revision 0.0 Date 10/15/09 Ballast Monitor Instruction Manual Changes made Original Release. Rev. 0.