Instruction Manual
Model TP-524D-HRT
TP-524D-HRT Instruction Manual Rev. 0.3 Page 44 of 48
9. Appendix
9.1 Specifications
Sensor Type: Continuous diffusion/adsorption type
CHEMFET Solid State MOS type
Sensor Life: 5-10 years typical
Measuring Ranges: 0-20ppm, 0-50ppm, or 0-100ppm
Accuracy/ Repeatability: ± 2ppm or ±10% of applied gas (whichever is greater)
Response/Clearing Time: T50 < 30 seconds, T80 < 60 seconds
Outputs: Linear 4-20mA
HART
Electrical Classification: Explosion Proof
CSA and US (NRTL)
Class I, Division 1, Groups B, C, D (Tamb= -40°C to 75°C)
Class I, Zone 1, Group IIB+H2
FM Approvals
Class I, Division 1, Groups B, C, D (Tamb= -40ºC ≤ Ta ≤ 60ºC)
ATEX
EExd IIB+H2 T4 (Tamb= -40ºC ≤ Ta ≤ 60ºC)
Performance Approvals: ISA 92.0.01 Part I-1998;
H2S Performance Requirements
Safety Approvals:
C
CSA
US
CSA/NRTL (US OSHA Certified)
FM
ATEX
CW Marking
Ingress Protection: NEMA 4X, IP66
Applicable Standards CSA C22.2 No. 30-M1986
CSA C22.2 No. 142-M1987
UL No. 916
UL No.1203
FM 3810
FM 3600
FM 3615
FM 6340
Warranty: Electronics – 2 years
Sensor – 10 year conditional
Environmental Specifications
Operating Temperature: -40°C to +75°C (-40°F to +167°F)