Storage Temperature: -35°C to +55°C (-31°F to +131°F)
Operating Humidity: 5-100% RH (Non-condensing)
Operating Pressure: Ambient ± 10%
Air Velocity: 0-5 meters/second
Mechanical Specifications
Mechanical Specifications
Dimensions: 10”H (including Splashguard) x 6.1”W x 4.3”D
(254mmH x 155mmW x 109mmD)
3 each ¾”NPT Female Input Ports
¼” Mounting Holes (J-box) 5.5” (140mm) Center to Center.
Weight: 6 lbs. (2.72kg) (with aluminum j-box)
9 lbs. (4.08kg (with Stainless Steel j-box)
Electrical Specifications
Electrical Specifications
Input Voltage: 12-28 VDC
Power Consumption: Normal operation = 68mA (<1.7 watt); Maximum = 85mA (2 watts)
RFI/EMI Protection: Complies with ISO 92.0.01
Analog Output: Linear 4-20mA DC current (1000 ohms maximum loop load @ 24VDC)
0mA All Fault Diagnostics
2mA In-Calibration
4-20mA 0-100% full-scale
22mA Over-range condition
HART Version 7.0
Manufacture ID 0x6007
Device ID 0xE08B
Status Indicators: 4-digit LED Display with gas concentration,
full-script menu prompts for AutoSpan,
Set-up Options, and Fault Reporting
LEDs for Fault and Calibration
Faults Monitored: Heater Fault, Loop Fault, Input Voltage Fault, Sensor Fault,
Processor Fault, Memory Fault, Calibration Fault(s)
Cable Requirements: Power/Analog – 3-wire shielded cable
Maximum distance is 13,300 feet with 14 AWG