User's Instructions
9-7
Symptom (continued)
Possible Cause
(continued)
Corrective Action (continued)
System returns to
Standby and reports
error message: “Vacuum
Error, Check Tubing
Connections” Error
Code 400 or 402.
Vacuum canister leak, or
vacuum pump is not
working properly.
• Ensure the vacuum tubing is connected as
defined in Chapter 3.
• Check that all vacuum canister lid ports are
sealed.
• Repeat system initialization.
• If the problem persists, call Customer
Service and report the error code displayed.
ST Only:
System instructs the
user to reinitialize the
holster: Error Code 904.
Holster firing fork has
encountered an obstruction
or there is an arming motor
failure.
• Check for and remove any obstructions that
may impede the holster firing fork.
• Reinitialize the holster.
• If the problem persists, call Customer
Service and report the error code displayed.
System reports: “Check
Holster and Probe
Attachment” during
function execution.
Error Codes 805 – 819.
System is unable to move
the cutter due to an
obstruction or probe issue.
• Remove the probe from the patient (collect
any samples obtained), disconnect the
probe from the system, and replace the
probe.
• If the problem persists with a new probe,
call Customer Service and report the error
code displayed.
System reports “Check
or Replace Cup” during
function execution:
Error Codes 705 – 720.
System is unable to move
the specimen management
system due to excessive
tissue or coagulating blood.
• If the problem persists, replace the
specimen management system.
If the problem persists with a new specimen
management system, replace the probe
(collect any samples obtained from the
original probe before disposal).
• If the problem persists with a new probe,
call Customer Service and report the error
code displayed.