User's Instructions

9-7
Symptom (continued)
Possible Cause
(continued)
Corrective Action (continued)
System returns to
Standby and reports
error message: “Vacuum
Error, Check Tubing
Connections” Error
Code 400 or 402.
Vacuum canister leak, or
vacuum pump is not
working properly.
Ensure the vacuum tubing is connected as
defined in Chapter 3.
Check that all vacuum canister lid ports are
sealed.
Repeat system initialization.
If the problem persists, call Customer
Service and report the error code displayed.
ST Only:
System instructs the
user to reinitialize the
holster: Error Code 904.
Holster firing fork has
encountered an obstruction
or there is an arming motor
failure.
Check for and remove any obstructions that
may impede the holster firing fork.
Reinitialize the holster.
If the problem persists, call Customer
Service and report the error code displayed.
System reports: “Check
Holster and Probe
Attachment” during
function execution.
Error Codes 805 819.
System is unable to move
the cutter due to an
obstruction or probe issue.
Remove the probe from the patient (collect
any samples obtained), disconnect the
probe from the system, and replace the
probe.
If the problem persists with a new probe,
call Customer Service and report the error
code displayed.
System reports “Check
or Replace Cup” during
function execution:
Error Codes 705 720.
System is unable to move
the specimen management
system due to excessive
tissue or coagulating blood.
If the problem persists, replace the
specimen management system.
If the problem persists with a new specimen
management system, replace the probe
(collect any samples obtained from the
original probe before disposal).
If the problem persists with a new probe,
call Customer Service and report the error
code displayed.