User's Instructions
2-14
NOTE: The sample collection area can be manually rotated by holding the outside of the sample
collection area to move to next desired chamber.
• VACUUM (VAC) button. Activates vacuum aspiration and probe clearing.
o Press and hold to activate vacuum aspiration for the duration of the press. The cutter retracts and
vacuum is applied for up to 15 seconds.
o Press and immediately release to activate a short vacuum and saline pulse. If a specimen is not
transported into its chamber, use this function to bring the clogged tissue or fluid into that
chamber, rather than indexing to the next sequential chamber.
NOTE: Post-Biopsy activation will transfer the tissue sample from the viewing window to the
collection chamber.
Remote Footswitch
The remote footswitch is a non-sterile, reusable device with a power cord to transmit information to the
control module. It provides three activation buttons for clinical functions.
Figure 15. Remote Footswitch Illustration and Nomenclature
Remote Keypad
The remote keypad is a non-sterile, reusable device that uses a power cord to transmit information to the
control module. It features three finger-tip operable buttons for clinical functions.