Unpacking the Cart
The cart comes fully assembled.
Remove cart from packaging.
The cart comes assembled with the Stereotactic Holster Docking Station on the right side of the cart and
the Ultrasound Docking Station on the left side. To switch docking station locations, use a 4mm hex key
to remove and reattach each docking station.
Items included in the shipping box:
o Cart
o Two (2) Thumbscrews attached to the Cart
Unpacking the EX Holder (Cart Detachable Component)
The EX Holder comes assembled.
Remove EX Holder from packaging.
The EX Holder will need to be attached to the cart, replacing the holder (docking station) on the left side
of the cart. To switch docking stations, follow the instructions provided in the EX Holder box.
Items included in the shipping box:
o EX Holder (Holster Holder and Probe Holder)
o Four (4) Screws (with washers)
o 4mm Hex Key Tool
o Removal and Attachment Instructions
Placing the Control Module on the Cart
1. Place the Mammotome revolve Control Module on the cart, ensuring the feet of the control module are
positioned in the recesses of the cart top base.
2. Secure the control module to the cart by tightening the two thumbscrews attached to the cart. The
thumbscrews are positioned under the recess of the cart top base.
3. Thread the power cord through the center column of the cart.
NOTE: Before transporting the system, ensure the wheels are unlocked and that all cords and tubes are
clear of the cart wheels.
NOTE: Before transporting the system, ensure that the footswitch, keypad, and holster are located on the
designated position on the cart to ensure that nothing falls from the cart.
WARNING: To ensure adequate ventilation, position the control module at least 6" (15.24 cm)
from all walls. Do not block the exhaust fan at the rear of the control module. Do not
position the device where it is difficult to operate the mains disconnect device. The
power switch is used for mains disconnect.
NOTE: When removing the Control Module from the cart, unplug the power cord, untighten the two
thumbscrews, and lift the Control Module by the handle and beneath the vacuum connection slot.
NOTE: Use care in placing the Control Module on the cart to ensure accurate performance.