User's Instructions

3-3
Unpacking the Holster
Lift the holster from the box and remove the protective wrapping.
Items included in the shipping box:
o ST Holster with attached electrical cable
OR
o U/S Holster with attached electrical cable
OR
o EX Holster with attached electrical cable
WARNING: Before using the holster, inspect its electric cable for fraying or damage. Do not use the
instrument if the cable shows signs of damage or wear or if the enclosure is broken.
Unpacking the Remote Footswitch
Lift the device from the box and remove the protective wrapping.
Items included in the shipping box:
o Remote Footswitch with attached power cord.
Unpacking the Remote Keypad
Lift the device from the box and remove the protective wrapping.
Items included in the shipping box:
o Remote Keypad with attached power cord.
Setup
1. Powering on the Control Module
NOTE: If using the cart and control module as a unit, ensure they have been properly assembled and set
up before attempting to power on the control module.
WARNING: To prevent any possibility of electrical shock, do not use the Control Module if it shows any
signs of damage.
a. Ensure that the power to the control module is off (O) before making electrical connections.
NOTE: Verify that the outlet provides the correct voltage for the Control Module before making any
electrical connections.
b. Plug the power cord into the grounded electrical outlet at the back of the control module. Plug the
opposite end of the power cord into a grounded wall.
WARNING: To avoid the risk of electrical shock, this equipment must only be connected to a supply
mains with protective earth.
NOTE: The Potential Equalization Conductor (identified by an inverted triangle) on the rear of the
control module provides a direct connection between the electrical equipment and the
potential equalization busbar of the electrical installation. This Potential Equalization
Conductor is provided in accordance with requirements of section 8.6.7 of IEC 60601-1
Edition 3.1.