User's Instructions

3-12
g. Carefully place the holster/probe assembly on the cart’s U/S docking station until the tissue sampling
procedure is ready to be initiated.
WARNING: When placing or removing the holster/probe assembly to or from the docking station on
the cart, ensure no contact is made between sterile and non-sterile surfaces (e.g., needle
tip to cart surfaces or control module).
EX Probe Only
d. Grasp the holster (with or without sleeve attached) and move toward the middle of the probe body.
e. Align the front tab of Mammotome revolve EX Holster/Sleeve assembly with the front tab slot of
Mammotome revolve EX Probe and push forward.
NOTE: The holster cable should be aligned with the cable slot on the left side of the probe.
f. With the tab and tab slot aligned, rock the holster, pivoting on the tab, so that the back surface of the
holster engages with the latching mechanism of the probe. An audible click may be heard, indicating
the probe is attached to the holster.