User's Instructions

3-13
NOTE: Care should be taken to prevent the holster from being dropped during probe attachment.
Dropping the holster may damage the holster.
g. Carefully place the holster/probe assembly on the cart’s EX Holder (docking station) for the probe
until the tissue sampling procedure is ready to be in initiated.
i. Place the tip of the sleeve protector (tip protector) into the semi-circular opening;
ii. Align the sleeve protector tab to the slot in the EX holder;
iii. Rock the holster/probe assembly downward, pivoting on the sleeve protector, until the
holster/probe assembly is held by the holder.
WARNING: When placing or removing the holster/probe assembly to or from the docking station on
the cart, ensure no contact is made between sterile and non-sterile surfaces (e.g., Needle
tip to cart surfaces or Control Module).
9. Connecting the Probe Vacuum Tube Set to the Mammotome revolve Dual Vacuum-Assisted
Biopsy System
a. Connecting the Vacuum Tube Set to the Saline Bag
i. Remove the saline spike cover from the saline spike.
NOTE: Saline shall be used for the procedure for consistent tissue transport to collection area(s).
ii. Connect the tubing line with the saline connection spike to a saline bag using standard technique.
NOTE: While connecting, removing, or transporting the saline bag, ensure to avoid spillage of
saline onto the floor, control module, and in the keypad storage receptacle because