User's Instructions

3-15
c. Connecting the Vacuum Tube Set to the Control Module
i. Using sterile technique, remove the vacuum tube set from the probe package. To avoid damage,
do not flip the device into the sterile field.
Note: Prior to use, remove and discard the retainers that secure the vacuum lines in the
packaging.
ii. Align the valve cartridge with the vacuum connection slot on the front of the control module,
making sure the locking tab is facing up and the filters are above the tab.
iii. Push the valve cartridge into the control module until it locks into place. Pull back on the valve
cartridge to ensure that it is engaged.
NOTES: Follow tube set assembly notes, below:
PRECAUTION: Ensure that the connection tubes are not placed over the control
module screen.
Avoid pinching the vacuum lines during assembly to the control module.
Do not force. Take care to avoid using excessive force when attaching the tube
set.
If using the cart, ensure the wheels are locked so the cart remains stable during
use.
Once the vacuum tube set has been connected, do not disconnect until ready to
discard both the probe and the vacuum tube set.
WARNING: Ensure that the connection tubes are not placed over the docked
handpiece.
10. Holster/Probe Assembly Initialization
When the probe and vacuum tubing have been attached, the Touchscreen interface displays the
initialization prompt:
This prompt appears when the control module has been turned on and the system detects a probe is
attached to a holster. If the system is booted up with a holster and probe attached, the startup
process automatically proceeds to this screen.