User's Instructions

3-16
To initialize the system:
a. Touch the forward arrow on the touchscreen interface to proceed with initialization. When initialization
is complete, the Touchscreen interface instructs the user to select an operating side at the holster (ST
Only). For Ultrasound guided procedures (U/S and EX), the software goes directly from Initialization
to the Ready Screen.
NOTE: Avoid pressing multiple buttons at once.
PRECAUTION: Leave the protective sleeve on the probe and keep hands away from the device
during initialization to avoid injury.
PRECAUTION: Ensure that no connection tubes are placed over the control module to ensure safety,
sterility, and correct operation.
11. Select Side of Holster and Ready for Procedure (ST Only)
After the system proceeds through all initializations (cutter initialization, specimen management
system initialization, vacuum initialization), the touchscreen interface of the control module displays a
prompt advising the user to take action at the ST holster:
The system displays this screen when the system is initialized, prompting the user to select which
side of the ST holster will be used during the procedure. When this screen is displayed on the control
module, the ST holster’s LEDs slowly flash blue, prompting the user to touch any flashing button on
either side of the holster to identify the desired operating side.
To select the operating side of the ST holster:
a. Press and hold (~2 seconds) any flashing illuminated button on the desired operating side of the
holster to continue.
NOTE: The placement of the holster buttons on the right-hand side of the holster is opposite of those