User's Instructions

4-11
Disconnecting Input Devices
a. Remove the power connector for the remote footswitch or remote keypad from the control
module’s remote device connection port.
NOTE: Input devices can be disconnected from the control module at any time.
b. Clean and/or disinfect the input device for later use as described in Chapter 7: Cleaning and
Disinfection.
Disconnecting the Vacuum Canister
a. Disconnect the control module’s flexible vacuum source tube from the center port of the lid.
b. Close all ports at the top of the vacuum canister.
c. Discard the vacuum canister in the appropriate container after use.
Shutting down the Mammotome revolve Dual Vacuum-Assisted Biopsy System
1. To shut down the system, select the green POWER button on the front of the control module. A
prompt on the screen asks the user to confirm the intended shutdown.
2. Touch YES to proceed to shut down.
o When YES is selected, the green POWER button on the front of the control module turns amber,
indicating the system has shut down.
o Touch NO to return to the previous screen.
3. Turn the control module off at the back of the system (On/Off power switch).
4. Unplug the power cord from the grounded wall outlet.
5. Before disconnecting the holster’s electrical cable from the control module, ensure the probe has
been removed from the holster.
6. Disconnect the holster electrical cable from the control module.
7. Clean and disinfect the holster for later use as described in Chapter 7: Cleaning and Disinfection.
NOTE: The Mammotome revolve Dual Vacuum-Assisted Biopsy System should be shut down once
per day.