Shutting Down the Mammotome revolve Dual Vacuum-Assisted Biopsy System
1. To shut down the system, press the green power button on the front of the control module. A prompt
appears on the screen, asking you to confirm the intended shutdown.
2. Touch YES to proceed to shut down.
o If YES is selected, the green POWER button on the front of the control module turns an
amber color, indicating the system has shut down.
o Touch NO to return to the previous screen.
NOTE: If the last button pressed was BIOPSY and the system hasn’t been placed into Standby, the
aperture will open and transport the last tissue sample from the viewing window to the tissue
sample chamber aligned.
3. Turn the control module off at the back of the system (On/Off power switch).
4. Unplug the power cord from the grounded wall outlet.
5. Before disconnecting the holster’s electrical cable from the control module, ensure the probe has
been removed from the holster.
6. Disconnect the holster electrical cable from the control module.
7. Clean and disinfect the holster for later use, as described in Chapter 7: Cleaning and Disinfection.