User's Instructions

Cleaning the Control Module, Cart, and Remote Footswitch: Option 2
Using Process Chemical Super Sani-Cloth (Manufacturer: PDI):
1. Use the specified wipe to wipe all external surfaces, including crevices to remove any debris.
2. Use the specified wipe to thoroughly wipe all external surfaces including cracks and crevices,
ensuring that a wet surface is maintained for 2.5 minutes.
Note: Additional wipes may be used to maintain the 2.5 minute contact time.
3. Use additional wipe(s) to repeat steps 1-2, above.
4. Allow the Control Module, Cart, and Remote Footswitch to air dry.
5. Perform a visual check to ensure all visible contamination is removed from the surface. If
contamination is present, repeat steps 1-3 until no visible contamination is present.
Cleaning and Disinfection Instructions for the Holsters and Remote Keypad
Cleaning the ST or U/S Holster: Option 1
Less than 45 minutes after each use, manually clean the holster thoroughly using a pH-neutral enzymatic
detergent. Proceed with manual cleaning using the following procedure:
1. Prepare a pH-neutral detergent or pH-neutral enzymatic detergent according to the manufacturer’s
directions.
2. Use a soft, clean cloth lightly moistened with the cleaning solution to manually clean all surfaces
(including crevices) that are exposed when the disposable probe is attached to the holster. The gears
on the holster should not be wiped with detergent.
3. To rinse, wipe thoroughly using a soft, clean cloth lightly moistened with lukewarm (27°C 44°C) tap
water.
Note: To rinse the crevices, wipe thoroughly using a clean soft bristle, standard toothbrush
moistened with lukewarm (27˚C- 44˚C) tap water.
4. Dry with a clean, soft cloth. Do not expose the device to heat greater than 54°C.
5. Inspect the device for visible contamination. If contamination is present, repeat steps 1-5 until no
visible contamination is present.
Cleaning the ST or U/S Holster: Option 2
Using Process Chemical Super Sani-Cloth (Manufacturer: PDI):
Additional Cleaning Tool(s): Soft Bristled Nylon Brush (brush head approximate dimensions: width 9 mm,
length 40 mm, individual bristles 9 mm long)
1. Use the specified wipe to wipe all external surfaces including crevices.
2. Thoroughly scrub crevice areas with a clean, dry soft bristled nylon brush for a minimum of 1 minute
10 seconds.
3. Use the specified wipe to thoroughly wipe the device, including crevices, ensuring that a wet surface
is maintained for 2.5 minutes.