User's Instructions

8-6
Advance to Chamber 7/Reset Chambers
Advance to Chamber 7 / Reset Chambers
Advance to Chamber 7
The Advance to Chamber 7 function is used to
skip the remaining chambers in the first set, to
start acquiring tissue samples in the second set of
chambers (7-12) in the specimen management
system.
The top center circle of the graphic depiction on
the touchscreen represents the marker chamber.
Chamber status is represented by three colors:
o Purple: Attempted tissue sample
o Dark Grey: Skipped/Advanced
chamber
o Light Grey: Available chamber
NOTE: If the first set of chambers (1-6) is removed
during the procedure (e.g., for specimen radiograph),
the system can continue to perform clinical functions
and retrieve tissue samples into the second set of
chambers (7-12) if “Advance to Chamber 7” is
activated.
Reset Chambers (continued)
The Reset Chamber function resets the graphic
image of the specimen management system on
the Touchscreen interface.
The Reset Chamber function is used when a new
specimen management system is placed on the
probe. Selecting “Reset Chambers” prompts the
system to calibrate proper chamber location, as
well as resetting the graphic representation of the
number of samples attempted on the Touchscreen
interface.
The system cannot detect when a specimen
management system has been removed from or
installed on the probe. Therefore, it is necessary to
use the Reset Chambers button when a new
specimen management system is installed on the
probe.
Advance to Chamber 7
(Example below: Three samples have been attempted.
“Advance to Chamber 7” is available for activation.)
Reset Chambers (continued)
(Example below: Three samples were attempted into
chambers 1-3, then “Advance to Chamber 7was
activated to skip Chambers 4-6. The next sample will
deposit in Chamber 7. The “Reset Chambers” function
is now available for activation if a new Specimen
Management System is installed.)