User's Instructions

8-11
Biopsy Counter
This number is seen in the center of the SMS
graphic and represents the number of attempted
samples taken during the biopsy procedure for a
single probe. The number increases as each
sample is attempted for a single probe.
The biopsy counter automatically resets when a
probe is removed from the holster. The user may
manually clear the count by pressing and holding
the center of the SMS graphic for ~2 seconds.
After probe initialization, if the probe remains
attached to the holster and the user does not
manually clear the count, the count tracks the
number of times the BIOPSY function is executed,
even when a probe has been placed in standby
then returned to the procedure to continue
sampling.
NOTE: Before any samples are attempted, no number
show on the biopsy counter. If manually cleared after
sampling, the count clears to zero (0).
NOTE: This number tracks the number of times the
BIOPSY function is executed. It DOES NOT represent
the number of chambers in the SMS, or the number of
the active SMS chamber
Set Vacuum Level
On this Confirm Selection Screen, the user can
adjust the vacuum level of the system. The default
vacuum level at startup is represented by the
middle bar in the graphic. If ambient air pressure
is too low for the pump to create the default
vacuum, the graphic shows the highest attainable
vacuum level as the first or second bar in the
graphic.
System shutdown returns the vacuum level to the
default setting.
To adjust setting: First, touch the available
vacuum level bar desired. Then touch the Confirm
button shown on the screen.
Default Setting: Level 3, or the closest attainable
vacuum level.