Sensor Glucose Alarm and Alerts
223
Initial Default Alert:
Vibrates 3 times.
After 5 Minutes:
Vibrates/beeps 3 times every 5 minutes until
cleared.
Trend screen will continue to reflect Alert until
sensor glucose readings go above Alert level.
Initial Default Alert:
Vibrates 2 times.
After 5 Minutes:
Vibrates/beeps 2 times every 5 minutes until
cleared.
Trend screen will continue to reflect Alert until
sensor glucose readings go below Alert level.
Rise Rate/Fall Rate Alerts
What you see
Beeps and vibration