subjects ages 6-17 also provided venous blood for comparison to a laboratory
method, the Yellow Springs Instrument 2300 STAT Plus™ Glucose Analyzer.
This instrument is referred to as the “YSI.” In the Original Pediatric Study,
subjects’ glucose levels were not intentionally manipulated during this study;
in the Software 505 Pediatric Study, subjects ages 13-17 had their
glucose levels intentionally manipulated during the clinic session. Readings
from the System were reported every 5 minutes and paired with YSI values
collected every 15 minutes in order to characterize how well the System
readings agreed with laboratory standard BG results. The remainder of the
study took place at home, and the System performance was also paired with
the comparative meter results, referred to as the “SMBG.”