User's Manual

Dexcom G5x System User Guide
Technical Information
284
Table 1-A. System Agreement to YSI Within CGM Glucose
Ranges (Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
than
40/40%
YSI
Overall
Original
9152
71%
82%
92%
3%
Software
505
2263
86%
93%
98%
1%
40-60
Original
512
67%
78%
88%
6%
Software
505
120
89%
94%
98%
0%
61-80
Original
781
73%
85%
94%
2%
Software
505
226
91%
96%
99%
0%
81-180
Original
3853
67%
78%
91%
3%
Software
505
738
84%
92%
98%
1%
181-300
Original
2784
72%
84%
93%
4%
Software
505
798
86%
93%
98%
1%
301-350
Original
775
82%
91%
97%
2%
Software
505
229
86%
94%
98%
1%
351-400
Original
447
74%
84%
91%
5%
Software
505
152
80%
92%
97%
0%
1
CGM readings are within 40-400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor