User's Manual

Dexcom G5x System User Guide
Technical Information
285
Table 1-B. System Agreement to YSI Within CGM Glucose
Ranges (Pediatric)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-YSI
Percent
Within
15/15%
YSI
Percent
Within
20/20%
YSI
Percent
Within
30/30%
YSI
Percent
Greater
than
40/40%
YSI
Overall
Original
2922
55%
68%
85%
7%
Software
505
2262
81%
91%
96%
2%
40-60
Original
19
63%
74%
79%
21%
Software
505
86
54%
74%
91%
3%
61-80
Original
76
61%
82%
92%
4%
Software
505
142
77%
82%
90%
3%
81-180
Original
1155
56%
69%
84%
6%
Software
505
805
78%
88%
97%
1%
181-300
Original
1380
55%
68%
85%
7%
Software
505
957
89%
96%
99%
1%
301-350
Original
206
48%
62%
80%
11%
Software
505
209
81%
91%
94%
5%
351-400
Original
86
48%
61%
79%
12%
Software
505
63
64%
81%
83%
8%
1
CGM readings are within 40-400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).