Table 3-A. Concurrence of CGM Readings and YSI Values
(Original Adult Study)
CGM
(mg/dL)
YSI (mg/dL)
Row Percentage of Matched Pairs in Each CGM Glucose Range
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>
400
Number
of Paired
CGM-YSI
< 40
6%
48%
37%
7%
1%
0%
0%
0%
0%
0%
0%
155
40-60
4%
49%
36%
11%
1%
0%
0%
0%
0%
0%
0%
512
61-80
0%
22%
51%
24%
1%
0%
0%
0%
0%
0%
0%
781
81-120
0%
2%
17%
66%
13%
1%
0%
0%
0%
0%
0%
1706
121-160
0%
0%
1%
25%
60%
13%
2%
0%
0%
0%
0%
1492
161-200
0%
0%
0%
2%
28%
53%
16%
2%
0%
0%
0%
1240
201-250
0%
0%
0%
0%
3%
21%
51%
21%
3%
1%
0%
1181
251-300
0%
0%
0%
0%
0%
4%
19%
49%
24%
3%
0%
1018
301-350
0%
0%
0%
0%
0%
0%
3%
28%
51%
16%
1%
775
351-400
0%
0%
0%
0%
0%
0%
3%
10%
43%
38%
7%
447
> 400
0%
0%
0%
0%
0%
0%
1%
6%
21%
57%
15%
248