Table 5-C. Hypoglycemia Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages
2-5 Years)
Hypoglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed
Detection Rate
55
Original
3%
97%
57%
43%
Software
505
25%
75%
100%
0%
60
Original
11%
89%
62%
38%
Software
505
20%
80%
100%
0%
70
Original
29%
71%
77%
23%
Software
505
20%
80%
100%
0%
80
Original
35%
65%
85%
15%
Software
505
61%
39%
100%
0%
90
Original
51%
49%
89%
11%
Software
505
78%
22%
100%
0%
100
Original
64%
36%
91%
9%
Software
505
82%
18%
100%
0%
1
Both sets of study data are presented and are labeled as Original (SW10050) or