User's Manual

Dexcom G5x System User Guide
Technical Information
305
Table 6-B. Hyperglycemia Alert and Detection Rate Evaluation in
Reference to YSI 15 Minutes Before and After (Pediatric, Ages 6-
17 Years)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection
Rate
Hyperglycemia
Missed
Detection
Rate
120
Original
91%
9%
98%
2%
Software
505
98%
2%
99%
1%
140
Original
87%
13%
99%
1%
Software
505
97%
3%
98%
2%
180
Original
75%
25%
99%
1%
Software
505
94%
6%
98%
2%
200
Original
71%
29%
98%
2%
Software
505
94%
6%
97%
3%
220
Original
67%
33%
97%
3%
Software
505
93%
7%
96%
4%
240
Original
62%
38%
96%
4%
Software
505
88%
12%
94%
6%
300
Original
43%
57%
93%
7%
Software
505
69%
31%
84%
16%
1
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor