User's Manual

Dexcom G5x System User Guide
Technical Information
306
Table 6-C. Hyperglycemia Alert and Detection Rate Evaluation in
Reference to SMBG 30 Minutes Before and After (Pediatric, Ages
2-5 Years)
Hyperglycemia
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hyperglycemia
Detection Rate
Hyperglycemia
Missed
Detection Rate
120
Original
92%
8%
98%
2%
Software
505
97%
3%
99%
1%
140
Original
90%
10%
98%
2%
Software
505
98%
2%
100%
0%
180
Original
87%
13%
96%
4%
Software
505
99%
1%
93%
7%
200
Original
85%
15%
96%
4%
Software
505
98%
2%
93%
7%
220
Original
81%
19%
95%
5%
Software
505
100%
0%
97%
3%
240
Original
80%
20%
95%
5%
Software
505
99%
1%
98%
2%
300
Original
71%
29%
90%
10%
Software
505
95%
5%
96%
4%
1
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor