User's Manual

Dexcom G5x System User Guide
Technical Information
312
Table 8-B. Sensor Stability Relative to YSI (Accuracy Over Time
1
)
- (Pediatric, Ages 6-17 Years)
Day
of
Wear
Study
2
Number
of
Paired
CGM-
YSI
Mean
Absolute
Percent
Differenc
es
Median
Absolute
Percent
Differenc
es
Percen
t
Within
15/15%
YSI
Percen
t
Within
20/20%
YSI
Percen
t
Within
30/30%
YSI
Percen
t
Greater
than
40/40%
YSI
Day
1
Original
1016
21.2%
15.8%
48%
61%
78%
15%
Software
505
740
12.7%
8.5%
75%
83%
91%
4%
Day
4
Original
810
16.0%
13.9%
52%
66%
87%
3%
Software
505
795
8.1%
6.7%
89%
97%
100%
0%
Day
7
Original
1096
15.1%
11.3%
63%
76%
89%
4%
Software
505
727
10.4%
8.4%
80%
91%
98%
1%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor