Table 11-A. System Agreement to SMBG Within CGM Glucose
Ranges (Adult)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-
SMBG
Percent
Within
15/15%
SMBG
Percent
Within
20/20%
SMBG
Percent
Within
30/30%
SMBG
Percent
Greater
than
40/40%
SMBG
Overall
Original
7508
69%
81%
94%
2%
Software
505
2992
77%
87%
96%
1%
40-60
Original
731
75%
84%
92%
4%
Software
505
221
73%
80%
87%
7%
61-80
Original
968
78%
86%
95%
1%
Software
505
336
77%
85%
95%
1%
81-180
Original
3141
65%
78%
93%
2%
Software
505
1362
74%
85%
96%
1%
181-300
Original
1960
68%
81%
94%
3%
Software
505
826
80%
90%
97%
1%
301-350
Original
450
77%
88%
98%
1%
Software
505
161
83%
93%
99%
0%
351-400
Original
258
75%
85%
95%
2%
Software
505
86
90%
93%
98%
1%
1
CGM readings are within 40 to 400 mg/dL, inclusive.