User's Manual

Dexcom G5x System User Guide
Technical Information
321
Table 11-B. System Agreement to SMBG Within CGM Glucose
Ranges (Pediatric)
CGM
Glucose
Range
1
(mg/dL)
Study
2
Number
of Paired
CGM-
SMBG
Percent
Within
15/15%
SMBG
Percent
Within
20/20%
SMBG
Percent
Within
30/30%
SMBG
Percent
Greater
than
40/40%
SMBG
Overall
Original
16,318
64%
76%
89%
5%
Software 505
4264
73%
84%
94%
2%
40-60
Original
487
44%
55%
68%
19%
Software 505
240
54%
71%
86%
7%
61-80
Original
1340
59%
70%
85%
7%
Software 505
399
64%
76%
92%
2%
81-180
Original
7084
62%
74%
90%
5%
Software 505
1650
72%
84%
95%
2%
181-300
Original
5627
69%
80%
90%
5%
Software 505
1526
79%
89%
97%
2%
301-350
Original
1176
65%
77%
90%
4%
Software 505
319
72%
83%
94%
2%
351-400
Original
604
58%
72%
86%
6%
Software 505
130
69%
79%
86%
8%
1
CGM readings are within 40 to 400 mg/dL, inclusive.
2
Both sets of study data are presented and are labeled as Original (SW10050) or
Software 505 (SW10505).
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor