Transmitter Performance Characteristics
Parameter
Performance Characteristic
TX/RX Frequencies
2.402-2.480 GHz
Bandwidth
1.02 MHz
Maximum Output Power
1.0 mW EIRP
Modulation
Gaussian Frequency-Shift Keying
Data Rate
1 Mbps
Data Communication Range
20 feet
The Dexcom G5x CGM System is safe for use on U.S. commercial airlines.
The G5x System is an M-PED (Medical-Portable Electronic Device), which
meets the FAA RTCA /DO-160 edition G section 21, Category M. It can be
used on aircraft according to the directions provided by the operator of the
aircraft.
This device can withstand exposure to common electrostatic discharge (ESD)
and electromagnetic interference (EMI).
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The transmitter (P/N 9445-01) is intended for use in the electromagnetic
environment specified in the next table. The customer or the user of the
transmitter should ensure that it is used in such an environment.