User's Manual

Dexcom G5x System User Guide
Technical Information
328
Transmitter Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
Transmitter
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Receiver Product Specifications
Part Number
MT22719
Reading Frequency
Every 5 minutes
Dimensions
Length: 4.0 inches
Width: 1.8 inches
Thickness: 0.5 inches
TX/RX Frequencies
2.402-2.480 GHz
Bandwidth
1.22 MHz
Maximum Output Power
2.5 mW EIRP
Modulation
Gaussian Frequency-Shift Keying
Data Rate
1 Mbps