Calibration
83
Warning
Do:
During a sensor’s shelf life, store it between 36° F-86° F.
Why: Storing the sensor incorrectly might cause the sensor glucose
readings to be incorrect.
Never store sensors and/or sensor packages in the freezer.
Consequences: Missing a severe low or high glucose event.
Warning
Don’t:
Insert the sensor component of the system in a site other than
the belly/abdomen (ages 2 years and older) or the upper buttocks (ages
2 to 17 years).
Why: The placement and insertion of the sensor component of the
system is not approved for other sites.
Consequences: The system has not been tested in other areas and
may not work properly if inserted in other areas.
PDF compression, OCR, web optimization using a watermarked evaluation copy of CVISION PDFCompressor