User's Manual

Table Of Contents
Dexcom G5 Mobile System User Guide
Sensor Glucose Alarm and Alerts
209
Low Repeat/High Repeat
What you see
Beeps and vibration
Initial Default Alert:
None/Off.
After Setting Change:
Vibrates 3 times.
After 5 Minutes:
Vibrates/beeps 3 times every 5 minutes until cleared
or sensor glucose reading rises above Alert level.
Will re-alert if sensor glucose readings drop at or
below 55 mg/dL.
Initial Default Alert:
None/Off.
After Setting Change:
Vibrates 2 times.
After 5 Minutes:
Vibrates/beeps 2 times every 5 minutes until cleared
or sensor glucose reading drops below Alert level.
Signal Loss Alert
What you see Beeps and vibration